Development of toluidine red particle agglutination-based turbidimetric immunoassay for anticardiolipin antibody detection in syphilis

J Infect Chemother. 2024 Mar 7:S1341-321X(24)00077-1. doi: 10.1016/j.jiac.2024.03.002. Online ahead of print.

Abstract

Introduction: Serological tests of non-treponemal and treponemal types are the most frequently used for syphilis diagnosis. Nontreponemal tests are used to monitor disease activity. Toluidine red unheated serum test (TRUST), as one of nontreponemal tests, is generally applicable to hospitals at different levels. However, accurate judgment of TRUST results is inseparable from an experienced and accurate operator. To reduce current shortcomings of manual TRUST method, we attempted to convert the manual TRUST test into automatic TRUST test, that is, to determine the degree of aggregation of toluidine red particles by detecting the absorbance value of serum after reaction with toluidine red particles.

Methods: 50 μL of serum sample and 80 μL toluidine red particles were added to 96-well plate. Then, the 96-well plate was placed on a microplate reader at medium grade for 8 min to mix. Then, plasma reagin reacted with toluidine red particles and promoted the aggregation of toluidine red particles to form a large clot, which eventually caused a decrease in the absorbance at 540 nm.

Results: The results showed that the specificity of the automatic TRUST test was 100%, the sensitivity was 87%. And this method showed 93.5% correlation with manual TRUST test. The developed method is simple and involves less subjectivity in reading results, opening new avenues for syphilis diagnostic testing.

Conclusion: Turbidimetric immunoassay can avoid the shortcomings of subjective interpretation, time-consuming and manual operation of manual TRUST method, and is more suitable for large-scale screening in health examination.