Evaluation of agreement in macular vascular density measured by different protocols: 4.5 mm × 4.5 mm and 6 mm × 6 mm

Indian J Ophthalmol. 2024 Jun 1;72(6):864-868. doi: 10.4103/IJO.IJO_2187_23. Epub 2024 Mar 8.

Abstract

Purpose: To evaluate the agreement in the macular vascular density (MVD) measured by two protocols: 4.5 mm × 4.5 mm and 6 mm × 6 mm.

Design: Cross-sectional observational study.

Methods: Healthy volunteers between the age group of 19 and 39 years were recruited. Topcon DRI OCT Triton plus (Topcon Corporation, Tokyo, Japan) was used to acquire the fovea-centered optical coherence tomography angiography (OCTA) image using two protocols: 4.5 mm × 4.5 mm and 6 mm × 6 mm. MVD was measured by the manufacturer software in the superficial capillary plexus slab in five regions: central, superior, nasal, inferior, and temporal subfields of early treatment diabetic retinopathy study grid in each protocol.

Results: The study included 79 eyes of 40 healthy volunteers. The difference in the mean MVD between two protocols was 3.84% in right eye and 4.2% in left eye in central subfield, 0.93% in right eye and 1.13% in left eye in superior subfield, 0.06% in right eye and 1.45% in left eye in nasal subfield, 1.65% in right eye and 0.7% in left eye in inferior subfield, 0.4% in right eye and 0.54% in left eye in temporal subfield. The measurements were significantly higher in 6 mm × 6 mm in central subfield in both the eyes and in nasal field in the left eye. Whereas, the measurements were significantly higher in 4.5 mm × 4.5 mm in superior and inferior subfield in both the eyes and in temporal subfield in the left eye.

Conclusion: The protocols should not be used interchangeably and it is necessary to include recommendation of the field of view to measure MVD while standardizing OCTA reporting.

Publication types

  • Observational Study

MeSH terms

  • Adult
  • Cross-Sectional Studies
  • Female
  • Fluorescein Angiography* / methods
  • Fundus Oculi
  • Healthy Volunteers
  • Humans
  • Macula Lutea* / blood supply
  • Macula Lutea* / diagnostic imaging
  • Male
  • Microvascular Density
  • Reproducibility of Results
  • Retinal Vessels* / diagnostic imaging
  • Tomography, Optical Coherence* / methods
  • Young Adult