Effect of Camera Parallax Angle on the Accuracy of Static Contact Angle Measurements

Langmuir. 2024 Mar 12;40(10):5090-5097. doi: 10.1021/acs.langmuir.3c03684. Epub 2024 Feb 26.

Abstract

Measuring the contact angle at the solid/liquid/vapor triple point in sessile drop experiments is one of the most popular and simple ways to quantify the wettability of surfaces and determine the surface free energy. Despite decades of technical advancements in contact angle measurements, which allowed for improving the precision of sessile drop measurements below ±1°, an often overlooked source of experimental error in these measurements originates from the camera's parallax angle (PA) - the angle between the camera optical axis and the sample stage surface. Here, we quantified the systematic errors in the measurement of contact angles due to the acquisition of drop images at finite PA values by simulating sessile drop experiments in which synthetic drops were created using the Young-Laplace equation. The absolute contact angle error induced by imaging drops at nonzero PAs was found to increase as the true contact angle (TCA) deviates from 90° and resulted in an overestimation (underestimation) of the contact angle for drops having TCAs lower (higher) than 90°. The computed absolute contact angle error reaches values as high as -20° (+12.2°) for drops having a TCA of 175° (5°) when imaged with a PA of 10°, thus indicating the importance of considering the PA when accurately quantifying contact angles in sessile drop experiments. The shape and, by extension, volume of the sessile drop was also found to affect the magnitude of the absolute contact angle error as sessile drops with higher apex curvatures exhibited lower absolute error than those with lower curvatures at any given PA. The outcomes of this work provide guidelines for minimizing systematic errors in sessile drop measurements due to the collection of drop images at nonzero PAs.