The Ultrasound Hepato-Jugular Reflux: Measuring the Hepato-Jugular Reflux with Ultrasound with Comparison to Invasive Right Heart Catheterization

Am J Med. 2024 Feb 23:S0002-9343(24)00097-4. doi: 10.1016/j.amjmed.2024.02.019. Online ahead of print.

Abstract

Background: Ultrasound can overcome barriers to visualizing the internal jugular vein, allowing hepato-jugular reflux and jugular venous pressure measurement. We aimed to determine operating characteristics of the ultrasound hepato-jugular reflux and ultrasound jugular venous pressure predicting right atrial and pulmonary capillary occlusion pressures.

Methods: In a prospective observational cohort at three US academic hospitals the hepato-jugular reflux and jugular venous pressure were measured with ultrasound before right heart catheterization. Receiver operating curves, likelihood ratios, and regression models were utilized to compare the ultrasound hepato-jugular reflux and ultrasound jugular venous pressure to the right atrial and pulmonary capillary occlusion pressures.

Results: In 99 adults undergoing right heart catheterization, an ultrasound hepato-jugular reflux had a negative likelihood ratio of 0.4 if 0 cm and a positive likelihood ratio of 4.3 if ≥ 1.5 cm for predicting a pulmonary capillary occlusion pressure ≥ 15 mmHg. Regression modeling predicting pulmonary capillary occlusion pressure was not only improved by including the ultrasound hepato-jugular reflux (P < .001), it was the more impactful predictor compared with the ultrasound jugular venous pressure (adjusted odds ratio 2.6 vs 1.2). The ultrasound hepato-jugular reflux showed substantial agreement (kappa 0.76; 95% confidence interval, 0.30-1.21), with poor agreement for the ultrasound jugular venous pressure (kappa 0.11; 95% confidence interval, -0.37-0.58).

Conclusion: In patients undergoing right heart catheterization, the ultrasound hepato-jugular reflux is reproducible, has modest impact on the probability of a normal pulmonary capillary occlusion pressure when 0 cm, and more substantial impact on the probability of an elevated pulmonary capillary occlusion pressure when ≥ 1.5 cm.

Keywords: Congestive heart failure; Hepato-jugular reflux; Jugular venous pressure; Physical exam.