Optimization of the LLNL/CAMS gas-accepting ion source and 1 MV compact AMS for natural abundance radiocarbon analysis of CO2

Nucl Instrum Methods Phys Res B. 2022 Nov 1:530:1-7. doi: 10.1016/j.nimb.2022.08.012. Epub 2022 Sep 11.

Abstract

The Lawrence Livermore National Laboratory - Center for Accelerator Mass Spectrometry (LLNL/CAMS) 1 MV AMS system was converted from a biomedical AMS instrument to a natural abundance 14C spectrometer. The system is equipped with a gas-accepting hybrid ion source capable of measuring both solid (graphite) and gaseous (CO2) samples. Here we describe a series of experiments intended to establish and optimize 14CO2 measurement capabilities at natural abundance levels. A maximum instantaneous ionization efficiency of 8 % was achieved with 3 % CO2 in helium at a flow rate of approximately 220 μL/min (3.5 μg C/min). For modern materials (e.g., OX I) we measured an average of 240 ± 50 14C counts/μg C, equivalent to a total system efficiency of approximately 3 %. Experimental CO2 samples with F14C values ranging from 0.20 to 1.05 measured as both graphite and directly as CO2 gas produced equivalent values with an average offset of < 2σ.

Keywords: Accelerator mass spectrometry; CO2; Gas-accepting ion source; Radiocarbon.