Development of an UPLC-MS/MS method for quantitative analysis of abexinostat levels in rat plasma and application of pharmacokinetics

BMC Chem. 2024 Feb 20;18(1):37. doi: 10.1186/s13065-024-01144-z.

Abstract

Broad-spectrum histone deacetylase inhibitors (HDACi) have excellent anti-tumor effects, such as abexinostat, which was a novel oral HDACi that was widely used in clinical treatment. The purpose of this study was to establish a rapid and reliable method for the detection of abexinostat concentrations in rat plasma using ultra-performance liquid chromatography tandem mass spectrometry (UPLC-MS/MS). The mobile phase we used was acetonitrile and 0.1% formic acid, and the internal standard (IS) was givinostat. Selective reaction monitoring (SRM) was used for detection with ion transitions at m/z 397.93 → 200.19 for abexinostat and m/z 422.01 → 186.11 for givinostat, respectively. The intra-day and inter-day precision of abexinostat were less than 11.5% and the intra-day and inter-day accuracy ranged from - 10.7% to 9.7% using this method. During the analysis process, the stability of the test sample was reliable. In addition, the recovery and matrix effects of this method were within acceptable limits. Finally, the method presented in this paper enabled accurate and quick determination of abexinostat levels in rat plasma from the pharmacokinetic study following gavage at a dose of 8.0 mg/kg abexinostat.

Keywords: Abexinostat; Givinostat; Pharmacokinetics; UPLC-MS/MS.