Design and validation of a custom-made system to measure transepithelial electrical impedance in human corneas preserved in active storage machine

Int J Pharm X. 2024 Feb 9:7:100234. doi: 10.1016/j.ijpx.2024.100234. eCollection 2024 Jun.

Abstract

Corneal epithelial barrier represents one of the major limitations to ocular drug delivery and can be explored non-invasively through the evaluation of its electrical properties. Human corneas stored in active storage machine (ASM) could represent an interesting physiological model to explore transcorneal drug penetration. We designed a new system adapted to human corneas preserved in ASM to explore corneal epithelial barrier function ex-vivo. A bipolar set-up including Ag/AgCl electrodes adaptors to fit the corneal ASM and a dedicated software was designed and tested on freshly excised porcine corneas (n = 59) and human corneas stored 14 days in ASM (n = 6). Porcine corneas presented significant and proportional decrease in corneal impedance in response to increasing-size epithelial ulcerations and acute exposure to benzalkonium chloride (BAC) 0.01 and 0.05%. Human corneas stored 14 days in ASM presented a significant increase in corneal impedance associated with the restoration of a multi-layer epithelium and an enhanced expression of tight junctions markers zonula occludens 1, claudin 1 and occludin. These results support the relevance of the developed approach to pursue the exploration and development of human corneas stored in ASM as a physiological pharmacological model.

Keywords: Benzalkonium chloride; Corneal active storage machine; Corneal epithelial barrier; Corneal impedance.