Investigation of the Effect of 70 Potential Interferents on Measurement Results of Two Blood Glucose Monitoring Systems

J Diabetes Sci Technol. 2024 Feb 15:19322968241231294. doi: 10.1177/19322968241231294. Online ahead of print.

Abstract

Background: Reliable blood glucose (BG) measurements are important for people with diabetes to manage their therapy as well as in point-of-care testing (POCT) performed by health care professionals to monitor BG of patients or even to diagnose diabetes. Among other factors, endogenous and exogenous substances present in blood samples can impact the measurement results. To ensure and prove that blood glucose monitoring systems (BGMSs) are robust in terms of potential interferents, manufacturers have to perform extensive evaluations.

Method: An interference screening test was performed for three reagent system lots of a POCT system and of a BGMS for self-monitoring of BG. A paired-difference approach based on ISO 15197:2013 and CLSI guideline EP07 was used with venous whole blood samples at two different glucose concentrations. Seventy potential interferents expected to be common in people with diabetes were evaluated.

Results: The interference effects were determined as normalized biases between test samples and corresponding control samples. For 69 of the 70 investigated potential interferents, both systems met the predefined acceptance criteria, with the normalized biases falling within ±10 mg/dL or ±10% at glucose concentrations ≤100 mg/dL or >100 mg/dL, respectively, for each of the three evaluated reagent system lots.

Conclusions: The BGMS investigated in this study were found to be robust with respect to the 70 evaluated potential interferents. Interference effects were observed only for N-Acetyl-L-cysteine. Extensive evaluations of potential interfering factors can make an important contribution to ensure reliability of BGMS.

Keywords: CLSI EP07; CLSI EP37; ISO 15197; blood glucose measurement; interference testing.