A remote-controlled automatic chest compression device capable of moving compression position during CPR: A pilot study in a mannequin and a swine model of cardiac arrest

PLoS One. 2024 Jan 19;19(1):e0297057. doi: 10.1371/journal.pone.0297057. eCollection 2024.

Abstract

Background: Recently, we developed a chest compression device that can move the chest compression position without interruption during CPR and be remotely controlled to minimize rescuer exposure to infectious diseases. The purpose of this study was to compare its performance with conventional mechanical CPR device in a mannequin and a swine model of cardiac arrest.

Materials and methods: A prototype of a remote-controlled automatic chest compression device (ROSCER) that can change the chest compression position without interruption during CPR was developed, and its performance was compared with LUCAS 3 in a mannequin and a swine model of cardiac arrest. In a swine model of cardiac arrest, 16 male pigs were randomly assigned into the two groups, ROSCER CPR (n = 8) and LUCAS 3 CPR (n = 8), respectively. During 5 minutes of CPR, hemodynamic parameters including aortic pressure, right atrial pressure, coronary perfusion pressure, common carotid blood flow, and end-tidal carbon dioxide partial pressure were measured.

Results: In the compression performance test using a mannequin, compression depth, compression time, decompression time, and plateau time were almost equal between ROSCER and LUCAS 3. In a swine model of cardiac arrest, coronary perfusion pressure showed no difference between the two groups (p = 0.409). Systolic aortic pressure and carotid blood flow were higher in the LUCAS 3 group than in the ROSCER group during 5 minutes of CPR (p < 0.001, p = 0.008, respectively). End-tidal CO2 level of the ROSCER group was initially lower than that of the LUCAS 3 group, but was higher over time (p = 0.022). A Kaplan-Meier survival analysis for ROSC also showed no difference between the two groups (p = 0.46).

Conclusion: The prototype of a remote-controlled automated chest compression device can move the chest compression position without interruption during CPR. In a mannequin and a swine model of cardiac arrest, the device showed no inferior performance to a conventional mechanical CPR device.

MeSH terms

  • Animals
  • Cardiopulmonary Resuscitation*
  • Heart Arrest* / therapy
  • Hemodynamics
  • Male
  • Manikins
  • Pilot Projects
  • Pressure
  • Swine

Grants and funding

This research was supported by a grant of the Korea Health Technology R&D Project through the Korea Health Industry Development Institute (KHIDI), funded by the Ministry of Health & Welfare, Republic of Korea (grant number: HW20C2132). GJS received this fund. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.