Dual-energy computed tomography quality control: Initial experiences with a semi-automatic analysis tool

Phys Med. 2024 Feb:118:103211. doi: 10.1016/j.ejmp.2024.103211. Epub 2024 Jan 18.

Abstract

Purpose: A quality control (QC) system for dual-energy CT (DECT) was developed. The scope of the QC system was to monitor both the constancy of the CT images and the software used in calculating the DECT derived maps. Longitudinal analysis was based on a standard imaging protocol, a commercial multi-energy phantom, and a semi-automatic analysis tool.

Methods: The phantom consisted of an elliptical body section with round slots for interchangeable inserts. It was scanned with 90kVp/Sn150kVp, automatic tube current modulation, and 9.6 mGy CTDIvol. From the two conventional CT images, scanner manufacturer's software was used to provide virtual monoenergetic images at two different energies, effective atomic number (Zeff) maps, and iodine concentration maps. The images were analyzed using an open-source tool allowing user-selected statistics of interest. The means and standard deviations of the phantom background and the iodine, calcium, and water inserts were recorded. The QC tool is available at github.com/tomakela/dectqatool.

Results: The obtained results were generally highly consistent over time, except for the smaller diameter iodine inserts. A small change inZeff was observed after a DECT software update. The developed QC tool aided the analysis robustness: the segmentations were modifiable when needed, and small rotations or air bubbles in the water insert were easily corrected.

Conclusion: The developed QC system provided easy-to-use workflow for constancy measurements. A small deviation due to change in the post-processing was detected. The proposed imaging protocol and analysis steps, and the reported measurement variations can aid in determining action levels for DECT QC.

Keywords: Dual-energy computed tomography; Phantom; Quality control.

MeSH terms

  • Iodine*
  • Phantoms, Imaging
  • Quality Control
  • Tomography, X-Ray Computed* / methods
  • Water

Substances

  • Iodine
  • Water