Insights on trends in the management of metrological analysis of devices in blood and transfusion medicine services: Verification based on their fitness for purpose and suitability for their respective applications

Transfus Apher Sci. 2024 Feb;63(1):103874. doi: 10.1016/j.transci.2024.103874. Epub 2024 Jan 11.

Abstract

In the field of blood and transfusion medicine, European Directorate for the Quality of Medicines & HealthCare recommends practices for metrological devices. Monitoring and measurement resources, as well as their permitted metrological limits, must be considered in this regard. We must understand variables such as metrological domains, the International System of Units, metrological traceability, and metrological requirements and controls in the field of transfusions in order to understand these metrological devices. An approach to metrological analysis and subsequent actions was constructed using flowcharts. As part of this methodology, a metrological device was selected, along with regulatory and normative requirements. A calibration and a test were conducted to demonstrate the application of the figure based on current recommendations. Current recommendations are consistent with the suggested ROC approach. The metrological devices must be verified based on their suitability for their respective applications (fitness-for-purpose).

Keywords: Calibration; Devices; Metrology; Test; Transfusion.

Publication types

  • Review

MeSH terms

  • Calibration
  • Humans
  • Reference Standards
  • Transfusion Medicine*