Acoustic resolution photoacoustic Doppler flowmetry for assessment of patient rectal cancer blood perfusion

J Biomed Opt. 2024 Jan;29(Suppl 1):S11517. doi: 10.1117/1.JBO.29.S1.S11517. Epub 2024 Jan 13.

Abstract

Significance: Photoacoustic Doppler flowmetry offers quantitative blood perfusion information in addition to photoacoustic vascular contrast for rectal cancer assessment.

Aim: We aim to develop and validate a correlational Doppler flowmetry utilizing an acoustic resolution photoacoustic microscopy (AR-PAM) system for blood perfusion analysis.

Approach: To extract blood perfusion information, we implemented AR-PAM Doppler flowmetry consisting of signal filtering and conditioning, A-line correlation, and angle compensation. We developed flow phantoms and contrast agent to systemically investigate the flowmetry's efficacy in a series of phantom studies. The developed correlational Doppler flowmetry was applied to images collected during in vivo AR-PAM for post-treatment rectal cancer evaluation.

Results: The linearity and accuracy of the Doppler flow measurement system were validated in phantom studies. Imaging rectal cancer patients treated with chemoradiation demonstrated the feasibility of using correlational Doppler flowmetry to assess treatment response and distinguish residual cancer from cancer-free tumor bed tissue and normal rectal tissue.

Conclusions: A new correlational Doppler flowmetry was developed and validated through systematic phantom evaluations. The results of its application to in vivo patients suggest it could be a useful addition to photoacoustic endoscopy for post-treatment rectal cancer assessment.

Keywords: acoustic resolution photoacoustic microscopy; endoscopy; photoacoustic Doppler flowmetry; rectal cancer; ultrasound.