A method to determine methanol (MeOH) and isopropyl alcohol (IPA) was developed using static headspace sampling (HSS) with gas chromatography (GC) and gas chromatography/mass spectrometry (GC/MS). The optimized HSS-GC and HSS-GC/MS methods were validated according to the parameters. The limit of quantification (LOQ) values for MeOH analyzed by HSS-GC/MS and HSS-GC were 0.50-0.56 and 1.97 mg/L, and the corresponding values for IPA were 0.14-0.21 and 1.51 mg/L, respectively. Recoveries were determined by spiking sample matrices, such as vinegar and tea. MeOH and IPA showed excellent recoveries of 93.80-107.60% and 93.92-104.32% with corresponding precisions of 0.51-7.38% and 0.90-7.70%, respectively. Compared to solid-phase microextraction-GC/MS and HSS-GC, the HSS-GC/MS method showed outstanding values for limit of detection and LOQ with improved precision and accuracy. This HSS-GC/MS method was successfully used to monitor residual solvents in 100 food and functional food products.
Keywords: Food; Headspace; Isopropyl alcohol; Methanol; Residual solvent.
© The Korean Society of Food Science and Technology 2023. Springer Nature or its licensor (e.g. a society or other partner) holds exclusive rights to this article under a publishing agreement with the author(s) or other rightsholder(s); author self-archiving of the accepted manuscript version of this article is solely governed by the terms of such publishing agreement and applicable law.