Trace gas sensor based on a multi-pass-retro-reflection-enhanced differential Helmholtz photoacoustic cell and a power amplified diode laser

Opt Express. 2024 Jan 1;32(1):848-856. doi: 10.1364/OE.512104.

Abstract

A high-sensitive photoacoustic spectroscopy (PAS) sensor, which is based on a multi-pass-retro-reflection-enhanced differential Helmholtz photoacoustic cell (DHPAC) and a high power diode laser amplified by erbium-doped fiber amplifier (EDFA), is presented in this work for the first time. In order to improve the interaction length between the light and target gas, the incident light was reflected four times through a multi-pass-retro-reflection-cell constructed by two right-angle prisms. A 1.53 µm distributed feedback (DFB) diode laser was selected to excite photoacoustic signal. Moreover, its power was amplified by an EDFA to 1000 mW to improve the amplitude of photoacoustic signal. Acetylene (C2H2) was chosen as the target analysis to verify the reported sensor performance. Compared to double channel without multiple reflections, the 2f signal of double channel with four reflections was improved by 3.71 times. In addition, when the output optical power of EDFA was 1000 mW, the 2f signal has a 70.57-fold improvement compared with the multi-pass-retro-reflection-cell without EDFA. An Allan deviation analysis was carried out to evaluate the long-term stability of such PAS sensor. When the averaging time was 400 s, the minimum detection limit (MDL) of such PAS sensor was 14 ppb.