Aim: Accurate working length determination during root canal treatment is essential for achieving successful outcomes. This study aimed to evaluate the impact of embedding medium on the accuracy of iPex electronic apex locator (EAL).

Materials and methods: Sixty-one extracted single-rooted teeth were decoronated and coronally flared with Gates-Glidden burs. Actual canal length (ACL) was obtained by introducing a size 8 K-file until its tip reached the most coronal border of the apical foramen. This step was performed thrice and then averaged. Deducting 0.5 mm from the ACL provided the working length (WL). The teeth were randomly placed in plastic containers filled with freshly mixed alginate, gelatin, or saline, with the lip clip placed in the medium. The blinded operator obtained electronic measurements using iPex by advancing a K-file with a size compatible with the canal attached to the file clip and advanced until the 0.0 mark, then withdrawn to the 0.5 mark. This step was performed thrice and then averaged. Data were analyzed using ANOVA and Tukey's post hoc test, with significance level set at 5% (α = 0.05).

Results: The mean difference between WL and iPex length obtained in the gelatin model was significantly longer than the difference with mean iPex length in alginate (p = 0.005) and in saline (p < 0.001). There was no significant difference between iPex readings obtained in alginate and saline (p = 0.249).

Conclusion: The use of freshly mixed alginate or saline for ex vivo assessment of iPex is recommended, whereas the use of gelatin could increase the chances of readings longer than looked for.

Clinical significance: Identifying the optimum embedding medium for ex vivo testing of EALs permits the comparison and assessment of several factors affecting EALs' precision under standardized conditions. This helps in understanding EAL performance in vivo and in optimizing its clinical utilization.

Keywords: Apex locators; Electronic apex locator; Electronic working length; Embedding medium; Endodontics; iPex Working length determination..