Comparison of two swept-source optical coherence tomography devices, a Scheimpflug camera system and a ray-tracing aberrometer in the measurement of corneal power in patients with cataract

Graefes Arch Clin Exp Ophthalmol. 2024 May;262(5):1567-1578. doi: 10.1007/s00417-023-06348-y. Epub 2023 Dec 27.

Abstract

Purpose: To assess the differences and similarities in the corneal curvature obtained by two swept-source optical coherence tomography (SS-OCT) devices, Scheimpflug imaging system and one ray tracing aberrometer in patients with cataracts. Moreover, this study aimed to compare the differences in posterior corneal (PK), total corneal (TK) and true net power (TNP) measurements among the IOLMaster 700, CASIA2, and Pentacam.

Methods: A total of 200 eyes of 200 patients (116 female, 58%) were enrolled in this study, with a mean age of 65.9 ± 9.5 years. The flattest (Kf), steepest (Ks), and mean cornal powers (Km), J0, and J45 were obtained using two SS-OCT-based biometric devices, one rotating camera system and one ray-tracing aberrometer. The PK, TK and TNP values were also measured using these devices. To evaluate the differences and similarities between the devicves, the Friedman test, Pearson correlation coefficient (r), intraclass coefficient correlation (ICC) and Bland‒Altman plots with 95% limits of agreement (LoA) were used, and boxplots and stacked histograms were generated to describe the distributions of the data.

Results: There were no significant differences between the IOLMaster 700 and Pentacam for any of the keratometry values. Additionally, there were no significant differences between the IOLMaster 700 and iTrace in evaluating J0 and J45. Bland‒Altman plots revealed relatively wide LoA widths, almost larger than 1 diopter for the keratometry values and almost larger than 0.5 diopter for J0 and J45 values among the four devices. In terms of PK and TK values, significant differences and low ICCs were found among the three devices.

Conclusions: Although strong correlations and good agreement were found among the IOLMaster700, CASIA2, Pentacam and iTrace for Kf, Ks, Km and J0, J45, it seems that the measurements should not be used interchangeably because of the wide LoA widths and the presence of significant differences among the devices. Similarly, due to significant differences and low ICCs, the PK, TK and TNP values obtained by IOLMaster 700, CASIA2, and Pentacam should not be used interchangeably.

Keywords: Astigmatic vector analysis; CASIA2; Corneal curvature; IOLMaster 700; Pentacam; iTrace.

MeSH terms

  • Aged
  • Biometry
  • Cataract* / diagnosis
  • Cornea
  • Corneal Topography / methods
  • Female
  • Humans
  • Middle Aged
  • Prospective Studies
  • Reproducibility of Results
  • Tomography, Optical Coherence* / methods