Automated method for the solid phase extraction of tetracyclines in wastewater followed by fluorimetric determination

Talanta. 2024 Apr 1:270:125544. doi: 10.1016/j.talanta.2023.125544. Epub 2023 Dec 13.

Abstract

This work presents a multisyringe flow injection analysis (MSFIA) system for the automatic extraction and determination of tetracyclines in wastewater samples. The sample was adjusted with Na2EDTA buffer before solid-phase extraction with an Oasis HLB column used for the analyte preconcentration. The europium (Eu3+)-based and citrate-mediated method (using Tris-HCl buffer) was selected for the fluorimetric analysis (λexc/em = 400/612 nm). For fluorescence detection, a low-cost system consisting of an USB 2000 CCD detector and a 3D-printed support that holds a LED light source was used. Under optimized conditions, the proposed method provided low limits of detection (9.4 μg L-1) and quantification (31 μg L-1), and good values for intra-day (<4 %) and inter-day precisions (<6 %). Recoveries of spiked TCs in wastewater samples ranged from 87 to 106 %. The results of this work were in good agreement with the measurements obtained by liquid chromatography coupled to a fluorescence detector.

Keywords: Automation; Fluorimetry; MSFIA; Solid-phase extraction; Tetracyclines; Wastewater analysis.

MeSH terms

  • Anti-Bacterial Agents
  • Chromatography, High Pressure Liquid / methods
  • Chromatography, Liquid
  • Solid Phase Extraction / methods
  • Tetracyclines* / analysis
  • Wastewater*

Substances

  • Tetracyclines
  • Wastewater
  • Anti-Bacterial Agents