Electrode-Skin Impedance Model Parameter Estimation in the Frequency-Domain

Annu Int Conf IEEE Eng Med Biol Soc. 2023 Jul:2023:1-4. doi: 10.1109/EMBC40787.2023.10340097.

Abstract

This paper presents a method for identifying parameter values for a double parallel resistor/constant-phase-element model of the electrode-skin interface for individual silver and silver/silver chloride electrodes. The impedance of each electrode was measured in five from 1 Hz-10 kHz. Phase features of these data were used to guide initial estimates for parameter values which were refined using a least squares algorithm. Resultant model impedances were compared with experimental data across a typical biosignal bandwidth (1 Hz-500 Hz). The method was effective in estimating component values in most datasets, and resulted in a mean relative RMS error of 7 % (σ = 8.3%) across the biosignal bandwidth.Clinical relevance- This work establishes a feature-based method for finding component parameter estimates for an electrode contact impedance model.

MeSH terms

  • Algorithms
  • Electric Impedance
  • Electrodes
  • Silver*
  • Skin*

Substances

  • Silver