Validity of Accelerometer-Based Sensor System for Muscle Tightness Estimation through Vibration on the Upper Limb

Annu Int Conf IEEE Eng Med Biol Soc. 2023 Jul:2023:1-4. doi: 10.1109/EMBC40787.2023.10340051.

Abstract

Spasticity is a condition that profoundly impacts the ability to perform everyday tasks. However, its diagnosis requires trained physicians and subjective evaluations that may vary depending on the evaluator. Focal vibration of spastic muscles has been proposed as a non-invasive, pain-free alternative for spasticity modulation. We propose a system to estimate muscular tightness based on the propagation of elastic waves in the skin generated by focal vibration of the upper limb. The developed system generates focalized displacements on the biceps muscle at frequencies from 50 to 200 Hz, measures the vibration acceleration on the vibration source (input) and the distant location (output), and extracts features of ratios between input and output. The system was tested on 5 healthy volunteers while lifting 1.25 - 11.25 kg weights to increase muscle tone resembling spastic conditions, where the vibration frequency and weight were selected as explanatory variables. An increase in the ratio of the root mean squares proportional to the weight was found, validating the feasibility of the current approach to estimating muscle tightness.Clinical Relevance- This work presents the feasibility of a vibration-based system as an alternative method to objectively diagnose the degree of spasticity.

MeSH terms

  • Accelerometry
  • Humans
  • Muscle Spasticity* / diagnosis
  • Muscle Tonus
  • Muscle, Skeletal / physiology
  • Upper Extremity
  • Vibration*