Computational Fluid Dynamics of Stent-Mounted Neural Interfaces in an Idealized Cerebral Venous Sinus

Annu Int Conf IEEE Eng Med Biol Soc. 2023 Jul:2023:1-4. doi: 10.1109/EMBC40787.2023.10341099.

Abstract

Hemodynamic changes in stented blood vessels play a critical role in stent-associated complications. The majority of work on the hemodynamics of stented blood vessels has focused on coronary arteries but not cerebral venous sinuses. With the emergence of endovascular electrophysiology, there is a growing interest in stenting cerebral blood vessels. We investigated the hemodynamic impact of a stent-mounted neural interface inside the cerebral venous sinus. The stent was virtually implanted into an idealized superior sagittal sinus (SSS) model. Local venous blood flow was simulated. Results showed that blood flow was altered by the stent, generating recirculation and low wall shear stress (WSS) around the device. However, the effect of the electrodes on blood flow was not prominent due to their small size. This is an early exploration of the hemodynamics of a stent-mounted neural interface. Future work will shed light on the key factors that influence blood flow and stenting outcomes.Clinical Relevance-The study investigates blood flow through a stent-based electrode array inside the cerebral venous sinus. The hemodynamic impact of the stent can provide insight into neointimal growth and thrombus formation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Cerebral Veins*
  • Coronary Vessels
  • Hemodynamics
  • Hydrodynamics*
  • Stents