Two-center validation of Pilot Tone based cardiac triggering of a comprehensive cardiovascular magnetic resonance examination

Int J Cardiovasc Imaging. 2024 Feb;40(2):261-273. doi: 10.1007/s10554-023-03002-w. Epub 2023 Dec 12.

Abstract

The electrocardiogram (ECG) signal is prone to distortions from gradient and radiofrequency interference and the magnetohydrodynamic effect during cardiovascular magnetic resonance imaging (CMR). Although Pilot Tone Cardiac (PTC) triggering has the potential to overcome these limitations, effectiveness across various CMR techniques has yet to be established. To evaluate the performance of PTC triggering in a comprehensive CMR exam. Fifteen volunteers and 20 patients were recruited at two centers. ECG triggered images were collected for comparison in a subset of sequences. The PTC trigger accuracy was evaluated against ECG in cine acquisitions. Two experienced readers scored image quality in PTC-triggered cine, late gadolinium enhancement (LGE), and T1- and T2-weighted dark-blood turbo spin echo (DB-TSE) images. Quantitative cardiac function, flow, and parametric mapping values obtained using PTC and ECG triggered sequences were compared. Breath-held segmented cine used for trigger timing analysis was collected in 15 volunteers and 14 patients. PTC calibration failed in three volunteers and one patient; ECG trigger recording failed in one patient. Out of 1987 total heartbeats, three mismatched trigger PTC-ECG pairs were found. Image quality scores showed no significant difference between PTC and ECG triggering. There was no significant difference found in quantitative measurements in volunteers. In patients, the only significant difference was found in post-contrast T1 (p = 0.04). ICC showed moderate to excellent agreement in all measurements. PTC performance was equivalent to ECG in terms of triggering consistency, image quality, and quantitative image measurements across multiple CMR applications.

Keywords: Cardiac triggering; Cardiovascular MRI; Pilot Tone; Prospective triggering.

MeSH terms

  • Caffeine
  • Contrast Media*
  • Gadolinium*
  • Humans
  • Magnetic Resonance Imaging
  • Magnetic Resonance Imaging, Cine
  • Magnetic Resonance Spectroscopy
  • Predictive Value of Tests

Substances

  • Contrast Media
  • Gadolinium
  • Caffeine