Finite Element Analysis of Subperiosteal Implants in Edentulism-On the Basis of the MaI Implant® by Integra Implants®

Materials (Basel). 2023 Nov 30;16(23):7466. doi: 10.3390/ma16237466.

Abstract

The MaI Implants® method offers a modern treatment option for specific patients who lack sufficient bone for traditional screw-based implants. The aim of the article is to use Finite Element Analysis (FEA) to examine the behavior of a subperiosteal implant under actual conditions within the oral cavity and to assess the impact of various mechanical factors on the durability of the MaI Implants®. A strength analysis was conducted using Finite Element Analysis for two models. The first was a single subperiosteal implant, while the second was a model of an arch consisting of two single subperiosteal implants connected by a bar. Based on the obtained results, it can be observed that the increase in load from 100 N to 800 N leads to an increase in displacements throughout the implant. Changing the angle from 90 to 30 degrees resulted in a 576% increase in the average displacement value across all multi-units. Stresses in the multi-units range from 23.7 MPa to 268.5 MPa. The lack of proper stabilization of the implant has the greatest impact on the results of displacements. Such displacements are significant for the later positioning of the implant compared to the initial conditions.

Keywords: FEA of dental implants; Integra Implants®; Mai Implants®; custom implant; individual implant; subperiosteal implant.

Grants and funding