Biomechanical design optimization of proximal humerus locked plates: A review

Injury. 2024 Feb;55(2):111247. doi: 10.1016/j.injury.2023.111247. Epub 2023 Nov 28.

Abstract

Background: Proximal humerus locked plates (PHLPs) are widely used for fracture surgery. Yet, non-union, malunion, infection, avascular necrosis, screw cut-out (i.e., perforation), fixation failure, and re-operation occur. Most biomechanical investigators compare a specific PHLP configuration to other implants like non-locked plates, nails, wires, and arthroplasties. However, it is unknown whether the PHLP configuration is biomechanically optimal according to some well-known biomechanical criteria. Therefore, this is the first review of the systematic optimization of plate and/or screw design variables for improved PHLP biomechanical performance.

Methods: The PubMed website was searched for papers using the terms "proximal humerus" or "shoulder" plus "biomechanics/biomechanical" plus "locked/locking plates". PHLP papers were included if they were (a) optimization studies that systematically varied plate and screw variables to determine their influence on PHLP's biomechanical performance; (b) focused on plate and screw variables rather than augmentation techniques (i.e., extra implants, bone struts, or cement); (c) published after the year 2000 signaling the commercial availability of locked plate technology; and (d) written in English.

Results: The 41 eligible papers involved experimental testing and/or finite element modeling. Plate variables investigated by these papers were geometry, material, and/or position, while screw variables studied were number, distribution, angle, size, and/or threads. Numerical outcomes given by these papers included stiffness, strength, fracture motion, bone and implant stress, and/or the number of loading cycles to failure. But, no paper fully optimized any plate or screw variable for a PHLP by simultaneously applying four well-established biomechanical criteria: (a) allow controlled fracture motion for early callus generation; (b) reduce bone and implant stress below the material's ultimate stress to prevent failure; (c) maintain sufficient bone-plate interface stress to reduce bone resorption (i.e., stress shielding); and (d) increase the number of loading cycles before failure for a clinically beneficial lifespan (i.e., fatigue life). Finally, this review made suggestions for future work, identified clinical implications, and assessed the quality of the papers reviewed.

Conclusions: Applying biomechanical optimization criteria can assist biomedical engineers in designing or evaluating PHLPs, so orthopaedic surgeons can have superior PHLP constructs for clinical use.

Keywords: Biomechanics; Locked plates; Optimization; Proximal humerus; Review.

Publication types

  • Review

MeSH terms

  • Biomechanical Phenomena
  • Bone Plates
  • Bone Screws
  • Fracture Fixation, Internal* / methods
  • Humans
  • Humerus
  • Shoulder Fractures* / surgery