Off-the-Shelf Single-Fenestrated Endograft for Emergent Juxtarenal and Pararenal Abdominal Aortic Aneurysm

J Endovasc Ther. 2023 Dec 4:15266028231215976. doi: 10.1177/15266028231215976. Online ahead of print.

Abstract

Introduction: Endovascular solutions to emergent juxtarenal and pararenal abdominal aortic aneurysms (AAAs) are complicated. Endovascular aortic repair (EVAR) with in situ laser fenestration (ISLF) is promising but requires a period of visceral ischemia. With an off-the-shelf, single superior mesenteric artery (SMA)-fenestrated device mesenteric ischemia is avoided and renal ischemia decreased. The aim was to develop an optimized design of such an endograft suitable for >90% of juxtarenal and pararenal AAAs.

Methods: Single-center analysis on 44 consecutive preoperative CTs for previously elective fenestrated EVARs for juxtarenal and pararenal aneurysms. Anatomical characteristics were analyzed to define: (1) shortest aortic coverage above SMA fenestration to achieve ≥4 cm seal; (2) feasibility of a scallop for the celiac artery; (3) shortest distance between the SMA and lowest renal, to facilitate renal ISLF in a straight endograft; (4) distance from the lowest renal to the aortic bifurcation, to allow an overlapping zone >40 mm with a bifurcated stent graft; (5) aortic diameter in the sealing zone, for optimal proximal stent graft diameter with 10% to 30% oversizing; (6) the final design was then tested on individual level.

Results: (1) The stent graft needs to start 40 mm above the SMA fenestration to achieve a 4 cm sealing zone in >90% of cases. (2) A proximal sealing zone of 40 mm without a scallop covers 77% of celiac arteries. With an addition of a 20 mm deep, 20 mm wide scallop at 12:30, the stent graft still covers 27% of celiacs. This suggests that a scallop would not be practically feasible. (3) In >90% of cases, the lowest renal was <31 mm from the SMA, suggesting that the tapering should start 30 mm below the SMA. (4) The distance from the lowest renal to the aortic bifurcation ranged from 82 to 166 mm. This allows for a 20 mm tapering and 50 mm straight part in all cases. (5) The 5th and 95th percentile of the aortic diameter in the sealing zone was 22 and 31 mm, respectively. Thus, 2 different stent graft diameters (28 and 34 mm) would fit >90% of cases. (6) The final design was suitable in 91% cases.

Conclusions: Two sizes of a single-fenestrated aortic stent graft without scallop cover >90% of juxtarenal and pararenal anatomies.

Clinical impact: Emergent juxta- and pararenal aortic aneurysms is a difficult clinical scenario that continuously challenges physicians. An endovascular option is in situ laser fenestrated endografts. One risk with these is the complete visceral ischemia occurring before the fenestrations are completed. An off-the-shelf single-fenestrated stent graft facilitates the treatment by removing the ischemia time for the SMA and reducing the ischemia time for the celiac and renal arteries thus decreasing the risk of visceral ischemia complications.

Keywords: custom-made device; emergent procedure; endovascular aneurysm repair; in situ laser fenestration; juxtarenal aneurysm; pararenal aneurysm; visceral ischemia.