Intra- and inter-fraction breath-hold variations and margins for radiotherapy of abdominal targets

Phys Imaging Radiat Oncol. 2023 Nov 11:28:100509. doi: 10.1016/j.phro.2023.100509. eCollection 2023 Oct.

Abstract

Radiotherapy in expiration breath-hold (EBH) has the potential to reduce treatment volumes of abdominal targets compared to an internal target volume concept in free-breathing. The reproducibility of EBH and required safety margins were investigated to quantify this volumetric benefit. Pre- and post-treatment diaphragm position difference and the positioning variability were determined on computed tomography. Systematic and random errors for EBH position reproducibility and positioning variability were calculated, resulting in margins of 7 to 12 mm depending on the prescription isodose and fractionation. A reduced volume was shown for EBH for lesions with superior-inferior breathing motion above 4 to 8 mm.

Keywords: Abdomen; Expiration breath-hold; Margin calculation; Stereotactic body radiotherapy.