A novel track guidance ultrasound facilitates radial arterial cannulation: A prospective randomized controlled trial

J Vasc Access. 2023 Nov 23:11297298231209382. doi: 10.1177/11297298231209382. Online ahead of print.

Abstract

Background: Ultrasound guidance for radial arterial cannulation is currently considered a best practice approach despite its clear advantages over the blind and palpation technique, the success rate is related to several factors, including clinician's experience and technical ability. The study aimed to explore the use of a novel track guidance ultrasound that may increase the success rate of radial arterial cannulation.

Methods: A randomized controlled trial was conducted, in which 80 adults scheduled for elective surgery requiring radial arterial cannulation were recruited and randomly assigned to either the experimental group, which utilized novel track ultrasound guidance (group T, n = 40), or the control group, which utilized traditional ultrasound guidance (group U, n = 40). The novel track guidance ultrasound comprises a positioning track and a guided track. The radial artery could be positioned at the center of the positional track on the ultrasound image, and the direction and angle of needle are fixed and toward the center of the positioning track. The primary endpoint of the study was the first-pass cannulation success rate, while the secondary endpoints included the failure rate of cannulation, the number of radial artery punctures, the time of cannulation, and the incidence of hematoma.

Results: The success rate of cannulation at the first attempt in group T (35 of 40 (87.5%)) was significantly higher than that in group U (23 of 40 (57.5%); p = 0.003). Although seven patients in Group U (7 of 40 (17.5%)) experienced failed cannulation compared to one in Group T (1 of 40 (2.5%)), the difference in failure rate between the two groups did not reach statistical significance (p = 0.06).

Conclusions: The implementation of novel track ultrasound guidance has demonstrated a notable improvement in the success rates at the first attempt while reducing the frequency of punctures and cannulation times.

Keywords: Ultrasonography; cannulation; complications; punctures; radial artery.