Validation of a Multiscale Computational Model Using a Mock Circulatory Loop to Simulate Cardiogenic Shock

ASAIO J. 2023 Dec 1;69(12):e502-e512. doi: 10.1097/MAT.0000000000002062. Epub 2023 Nov 3.

Abstract

The objectives of this study are to characterize the hemodynamics of cardiogenic shock (CS) through a computational model validated using a mock circulatory loop (MCL) and to perform sensitivity analysis and uncertainty propagation studies after the American Society of Mechanical Engineers (ASME) Validation and Verification (V&V) guidelines. The uncertainties in cardiac cycle time ( ), total resistance ( ), and total volume ( ) were quantified in the MCL and propagated in the computational model. Both models were used to quantify the pressure in the left atrium, aorta (Ao), and left ventricle (LV), along with the flow through the aortic valve, reaching a good agreement. The results suggest that 1) is the main source of uncertainty in the variables under study, 2) showed its greatest impact on the uncertainty of Ao hemodynamics, and 3) mostly affected the uncertainty of LV pressure and Ao flow at the late-systolic phase. Comparison of uncertainty levels in the computational and experimental results was used to infer the presence of additional contributing factors that were not captured and propagated during a first analysis. Future work will expand upon this study to analyze the impact of mechanical circulatory support devices, such as ventricular assist devices, under CS conditions.

MeSH terms

  • Computer Simulation
  • Heart Ventricles
  • Heart-Assist Devices*
  • Hemodynamics
  • Humans
  • Shock, Cardiogenic*