Performance and Safety of a Reflux-Control Microcatheter Used to Perform DEB-TACE with LUMITM Beads in HCC Patients: Preliminary Experience

J Clin Med. 2023 Oct 19;12(20):6630. doi: 10.3390/jcm12206630.

Abstract

Purpose: The present study aims to evaluate the effectiveness and safety of the anti-reflux microcatheter during DEB-TACE with DC Bead LUMITM (radiopaque beads) for the treatment of hepatocellular carcinoma (HCC).

Methods: We performed an observational longitudinal prospective monocentric study to analyze all patients with HCC who underwent to DEB-TACE with DC Bead LUMITM and anti-reflux microcatheter. Technical success, the presence of residual disease, and clinical success were evaluated. The performance of the anti-reflux microcatheter on the basis of the percentage of tumor covered and the non-target embolization (NTE) was also evaluated.

Results: Twenty patients underwent DEB-TACE with DC Bead LUMITM and an anti-reflux microcatheter. Technical success was achieved in all cases. Residual disease in the target tumor was observed in 11/20 (55%) of cases and no residual disease was found in 9/20 (45%) of cases. The clinical response at 1-month follow-up was of PD 4/20 (20%), SD 7/20 (35%), and CR 9/20 (45%). No major complications were recorded, and 10% of cases had minor complications. The distribution of beads on post-procedural CBCT, classified according to the percentage of target nodule coverage, was ≥50% in 70% (14/20) of cases and between 30-50% in 30% of cases (6/20). NTE was never registered.

Keywords: DEB-TACE; anti-reflux microcatheter; hepatocellular carcinoma (HCC); radiopaque microspheres; the non-target embolization (NTE).