Comparison of Safety of Leadless Pacemakers and Transvenous Pacemakers: A Meta-Analysis

Cureus. 2023 Sep 12;15(9):e45086. doi: 10.7759/cureus.45086. eCollection 2023 Sep.

Abstract

Pacemakers have been accessible for six decades, and clearly defined criteria for pacemaker implantation have been established. Within the contemporary clinical practice, two dependable pacing platforms exist leadless pacemakers and transvenous pacemakers. The aim of this meta-analysis is to compare the safety of leadless pacemakers to transvenous pacemakers. This meta-analysis adhered to the guidelines outlined in the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) 2020 framework. A comprehensive and systematic search was conducted across various databases including Scopus, Cochrane Library, and EMBASE, spanning from inception to August 15, 2023. The primary outcomes assessed in this meta-analysis were total complications, all-cause mortality, and device-related complications. Furthermore, secondary outcomes evaluated encompassed the need for reintervention, occurrences of pneumothorax, pericardial effusion, endocarditis, hemothorax, and hematoma. Total 17 studies were included in this meta-analysis. The findings of this study showed that patients with leadless pacemakers had a lower risk of total complications, device-related complications, pneumothorax, and endocarditis. The risk of reintervention was significantly lower in the leadless pacemaker group. However, compared to a transvenous pacemaker, the risk of pericardial effusion was significantly higher in the leadless pacemaker group. It is important to acknowledge the limitations arising from the lack of extensive long-term follow-up data for leadless pacemakers. As technology evolves, continued research will be essential in uncovering the full spectrum of prolonged complications associated with these devices.

Keywords: complications; leadless pacemakers; meta-analysis; safety; transvenous pacemakers.

Publication types

  • Review