Comparison of biological weighting functions to estimate the microdosimetric RBE in BNCT

Radiat Prot Dosimetry. 2023 Oct 11;199(15-16):1963-1967. doi: 10.1093/rpd/ncad007.

Abstract

In the framework of the MUNES project, a prototype accelerator-based thermal neutron source was developed and installed at the Legnaro National Laboratories of INFN. The microdosimetric characterization of this radiation field was performed with a Tissue-Equivalent Proportional Counter with interchangeable cathode walls, either doped with 100 ppm of 10B or without boron doping. A suitable subtraction procedure allowed to discriminate the gamma, neutron and BNC dose components (Selva et al., 2022, Appl. Radiat. Isot. 182, 110144). The measured microdosimetric spectra can be weighted with a biological weighting function to estimate the Relative Biological Effectiveness of the radiation field, for the purpose of intercomparison between different thermal neutron sources. This work compares, therefore, the biological doses resulting from four different weighting functions applied to the same initial microdosimetric spectrum, discussing strengths and limitations of each of them.

MeSH terms

  • Boron
  • Boron Neutron Capture Therapy* / methods
  • Gamma Rays
  • Neutrons
  • Radiation Dosage*
  • Radiation Monitoring*
  • Radiometry
  • Relative Biological Effectiveness*