Self-rectifying magnetoelectric metamaterials for remote neural stimulation and motor function restoration

Nat Mater. 2024 Jan;23(1):139-146. doi: 10.1038/s41563-023-01680-4. Epub 2023 Oct 9.

Abstract

Magnetoelectric materials convert magnetic fields into electric fields. These materials are often used in wireless electronic and biomedical applications. For example, magnetoelectrics could enable the remote stimulation of neural tissue, but the optimal resonance frequencies are typically too high to stimulate neural activity. Here we describe a self-rectifying magnetoelectric metamaterial for a precisely timed neural stimulation. This metamaterial relies on nonlinear charge transport across semiconductor layers that allow the material to generate a steady bias voltage in the presence of an alternating magnetic field. We generate arbitrary pulse sequences with time-averaged voltage biases in excess of 2 V. As a result, we can use magnetoelectric nonlinear metamaterials to wirelessly stimulate peripheral nerves to restore a sensory reflex in an anaesthetized rat model and restore signal propagation in a severed nerve with latencies of less than 5 ms. Overall, these results showing the rational design of magnetoelectric metamaterials support applications in advanced biotechnology and electronics.