Risk of Monopolar Electrosurgery in Cochlear Implant Recipients is Nominal: Evidence to Guide Clinical Practice

Otolaryngol Head Neck Surg. 2024 Feb;170(2):505-514. doi: 10.1002/ohn.555. Epub 2023 Oct 9.

Abstract

Objective: Comprehensively assess the prevalence of monopolar electrosurgery-related device complications among cochlear implant (CI) recipients.

Study design: Multifaceted retrospective review and survey.

Setting: Tertiary medical center.

Methods: Multifaceted approach including: (i) review of the current literature; (ii) historical review of institutional data from an academic, tertiary CI center; (iii) review of industry data provided by 3 Food and Drug Administration-approved CI manufacturers; and (iv) survey of high-volume CI centers.

Results: Literature review identified 9 human studies, detailing 84 devices with 199 episodes of device-cautery exposure. From studies reporting on patients records, no implant showed evidence of damage after exposure. One cadaveric study using dental cautery reported 1 episode of device damage. Review of institutional records did not identify any CI damage in 84 instances of exposure. Data from the 3 major implant manufacturers showed a single report of damage that could be reasonably linked to monopolar electrosurgery, out of a possible 689,426 CIs. Last, a survey of 8 high-volume CI centers did not identify any adverse events associated with monopolar cautery.

Conclusion: These data estimate the risk of adverse device-related events or tissue injury to be extraordinarily low. Short of operating in immediate proximity to the CI (ie, the ipsilateral temporoparietal scalp), these data indicate that monopolar electrosurgery can be used in the body and the head-and-neck of CI recipients with nominal risk. These findings may guide decision-making in cases that are optimally or preferably performed with monopolar electrocautery and can be used to counsel CI patients following inadvertent exposures.

Keywords: Bovie; cochlear implant; electrosurgery; monopolar cautery.

MeSH terms

  • Cautery
  • Cochlear Implantation* / adverse effects
  • Cochlear Implants* / adverse effects
  • Electrocoagulation
  • Electrosurgery / adverse effects
  • Humans