Needle guidance with Doppler-tracked polarization-sensitive optical coherence tomography

J Biomed Opt. 2023 Oct;28(10):102910. doi: 10.1117/1.JBO.28.10.102910. Epub 2023 Oct 4.

Abstract

Significance: Optical coherence tomography (OCT) can be integrated into needle probes to provide real-time navigational guidance. However, unscanned implementations, which are the simplest to build, often struggle to discriminate the relevant tissues.

Aim: We explore the use of polarization-sensitive (PS) methods as a means to enhance signal interpretability within unscanned coherence tomography probes.

Approach: Broadband light from a laser centered at 1310 nm was sent through a fiber that was embedded into a needle. The polarization signal from OCT fringes was combined with Doppler-based tracking to create visualizations of the birefringence properties of the tissue. Experiments were performed in (i) well-understood structured tissues (salmon and shrimp) and (ii) ex vivo porcine spine. The porcine experiments were selected to illustrate an epidural guidance use case.

Results: In the porcine spine, unscanned and Doppler-tracked PS OCT imaging data successfully identified the skin, subcutaneous tissue, ligament, and epidural spaces during needle insertion.

Conclusions: PS imaging within a needle probe improves signal interpretability relative to structural OCT methods and may advance the clinical utility of unscanned OCT needle probes in a variety of applications.

Keywords: birefringence; imaging coherence; lasers in medicine; polarization.

Publication types

  • Research Support, U.S. Gov't, Non-P.H.S.
  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Birefringence
  • Light
  • Refraction, Ocular
  • Skin* / diagnostic imaging
  • Swine
  • Tomography, Optical Coherence* / methods