The Effect of Bone Quality on Treatment of Intertrochanteric Fractures with Helical Blade Versus Lag Screw Fixation in Cephalomedulary Nails

Kans J Med. 2023 Aug 24:16:207-213. doi: 10.17161/kjm.vol16.20105. eCollection 2023.

Abstract

Introduction: The specific aim of this retrospective study was to determine whether bone quality has any effect on the complication rates or overall survivorship between helical blades and lag screws in cephalomedullary nails used for intertrochanteric hip fractures.

Methods: The authors reviewed clinical charts and radiographic studies of patients between January 2012 and August 2019. We reviewed radiographic images (pre-, intra-, and post-operative) to evaluate fracture fixation type, fracture reduction grade, and post-operative complications. We collected dual energy x-ray absorptiometry scan results (T-score) and serum alkaline phosphatase (ALP) isoenzyme activity values to evaluate patient bone quality.

Results: We included 303 cases (helical: 197, screw: 106) in the study. Complications were found in 31 (16%) helical blade cases and 23 (22%) lag screw cases. No statistically significant difference was detected when comparing complication rates with patient bone quality between the two groups. These two groups had similar one-year implant survivorship with respect to T-score, the low ALP level group, and normal ALP level group. The helical blade had higher implant survivorship compared to lag screw in five-year survival rate with respect to osteoporotic group, high ALP level group, and normal ALP level group (osteoporotic: 77% vs 69%, high ALP: 73% vs 67%, normal ALP: 70% vs 64%).

Conclusions: Similar complication rates were observed between helical blade and lag screw constructs in cephalomedullary femoral nails when accounting for patient bone quality. However, the helical blade design had a higher five-year survival rate.

Keywords: helical blade screw; hip fractures; intertrochanteric fractures; intramedullary nailing; lag screw.