Physics-Inspired Regularized Pulse-Echo Quantitative Ultrasound: Efficient Optimization With ADMM

IEEE Trans Ultrason Ferroelectr Freq Control. 2023 Nov;70(11):1428-1441. doi: 10.1109/TUFFC.2023.3321250. Epub 2023 Nov 1.

Abstract

Pulse-echo quantitative ultrasound (PEQUS), which estimates the quantitative properties of tissue microstructure, entails estimating the average attenuation and the backscatter coefficient (BSC). Growing recent research has focused on the regularized estimation of these parameters. Herein, we make two contributions to this field: first, we consider the physics of the average attenuation and backscattering to devise regularization terms accordingly. More specifically, since the average attenuation gradually alters in different parts of the tissue, while BSC can vary markedly from tissue to tissue, we apply L2 and L1 norms for the average attenuation and the BSC, respectively. Second, we multiply different frequencies and depths of the power spectra with different weights according to their noise levels. Our rationale is that the high-frequency contents of the power spectra at deep regions have a low signal-to-noise ratio (SNR). We exploit the alternating direction method of multipliers (ADMM) for optimizing the cost function. The qualitative and quantitative evaluations of bias and variance exhibit that our proposed algorithm improves the estimations of the average attenuation and the BSC up to about 100%.