A Randomised Clinical Feasibility Trial of a Breast Immobilisation Device: The SuPPORT 4 All Bra

Clin Oncol (R Coll Radiol). 2023 Dec;35(12):801-810. doi: 10.1016/j.clon.2023.09.008. Epub 2023 Sep 22.

Abstract

Aims: Despite the breast being a mobile organ, there is currently no standard suitable immobilisation device to optimise radiotherapy for women with larger breasts treated after a wide local excision. The SuPPORT 4 All (S4A) bra was co-designed with patients and radiotherapy professionals. The purpose of this study was to test the feasibility of using the S4A bra in the existing breast cancer radiotherapy pathway.

Materials and methods: A randomised feasibility trial was conducted in a single institution; the primary feasibility endpoint was the recruitment of 50 participants. Efficacy endpoints were also tested, including assessment of skin reactions, dose to organs at risk and patient comfort. Fifty women were randomised to receive either standard radiotherapy with no immobilisation (control) or radiotherapy with the S4A bra (intervention). A separate planning study was undertaken on the cases randomised to receive the S4A bra. Participants in the intervention arm (S4A bra) underwent two planning computed tomography scans, one with the bra on and one without the bra; allowing direct comparison of organs at risk data for S4A bra versus no bra.

Results: All women who started radiotherapy wearing the S4A bra completed treatment with the bra; patient comfort did not change across the 3 weeks of treatment. Positional accuracy using the bra was comparable with existing published accuracy for methods without immobilisation. The mean ipsilateral lung doses showed some improvement when positioning with the S4A bra was compared with the no bra set-up (3.72 Gy versus 4.85 Gy for right-sided cases, 3.23 Gy versus 3.62 Gy for left-sided cases).

Conclusions: The S4A bra is feasible to use in the radiotherapy pathway with good patient adherence. The S4A bra has potential to reduce dose to organs at risk (specifically ipsilateral lung dose) while maintaining good breast tissue coverage, and improved patient dignity, warranting further investigation on a larger scale.

Keywords: Breast radiotherapy; immobilisation device; support bra.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Breast Neoplasms* / radiotherapy
  • Breast*
  • Feasibility Studies
  • Female
  • Humans
  • Lung
  • Radiotherapy Dosage
  • Radiotherapy Planning, Computer-Assisted / methods