Predicting mammographic density with linear ultrasound transducers

Eur J Med Res. 2023 Sep 28;28(1):384. doi: 10.1186/s40001-023-01327-9.

Abstract

Background: High mammographic density (MD) is a risk factor for the development of breast cancer (BC). Changes in MD are influenced by multiple factors such as age, BMI, number of full-term pregnancies and lactating periods. To learn more about MD, it is important to establish non-radiation-based, alternative examination methods to mammography such as ultrasound assessments.

Methods: We analyzed data from 168 patients who underwent standard-of-care mammography and performed additional ultrasound assessment of the breast using a high-frequency (12 MHz) linear probe of the VOLUSON® 730 Expert system (GE Medical Systems Kretztechnik GmbH & Co OHG, Austria). Gray level bins were calculated from ultrasound images to characterize mammographic density. Percentage mammographic density (PMD) was predicted by gray level bins using various regression models.

Results: Gray level bins and PMD correlated to a certain extent. Spearman's ρ ranged from - 0.18 to 0.32. The random forest model turned out to be the most accurate prediction model (cross-validated R2, 0.255). Overall, ultrasound images from the VOLUSON® 730 Expert device in this study showed limited predictive power for PMD when correlated with the corresponding mammograms.

Conclusions: In our present work, no reliable prediction of PMD using ultrasound imaging could be observed. As previous studies showed a reasonable correlation, predictive power seems to be highly dependent on the device used. Identifying feasible non-radiation imaging methods of the breast and their predictive power remains an important topic and warrants further evaluation. Trial registration 325-19 B (Ethics Committee of the medical faculty at Friedrich Alexander University of Erlangen-Nuremberg, Erlangen, Germany).

Keywords: Breast cancer risk; Percent mammographic density; Ultrasound.

MeSH terms

  • Breast Density*
  • Breast Neoplasms* / diagnostic imaging
  • Female
  • Humans
  • Lactation
  • Mammography / methods
  • Pregnancy
  • Risk Factors
  • Transducers