Different Methods for Assessing Tooth Colour-In Vitro Study

Biomimetics (Basel). 2023 Aug 23;8(5):384. doi: 10.3390/biomimetics8050384.

Abstract

Colour assessment using digital methods can yield varying results, and it is important for clinicians to recognize the potential variability intra and inter-device. This study aimed to compare the L*a*b* values of VITA Classical (VC) and VITA Toothguide 3D-MASTER (VM) guides using two methods, SpectroShade (SS) and eLAB. Thirty-four measurements per tab were performed by a single operator across three batches of each guide. Intraclass correlation coefficients (ICC) between batches were calculated. Values <0.5, 0.5-0.75, 0.75-0.9, and >0.90 were classified as poor, moderate, good, and excellent reliability, respectively. Results were reported as mean and standard deviation of the L*a*b* values and respective colour differences (ΔE00) for each tab and method. Statistical analyses were performed with an independent t-test, α = 0.05. ICC values between batches were excellent for all L*a*b*, except for a* component in eLAB. There were statistically significant differences between methods in most L*a*b* values. The intra-device mean ΔE00 was 0.5 ± 0.6 for VC, 0.5 ± 0.8 for VM in SS, 1.1 ± 0.8 for VC, 1.1 ± 0.9 for VM in eLAB. The mean ΔE00 inter-device was 4.9 ± 1.7 for VC, 5.0 ± 1.7 for VM. Both methods demonstrated good internal consistency, with high ICC values and low intra-device colour differences, but exhibited high variability between methods, higher for a* the component.

Keywords: SpectroShade; eLAB; photography; spectrophotometry; tooth colour.