Continuous aerosol monitoring and comparison of aerosol exposure based on smoke dispersion distance and concentrations during oxygenation therapy

Sci Rep. 2023 Sep 23;13(1):15910. doi: 10.1038/s41598-023-42909-1.

Abstract

This study evaluated the aerosol exposure risks while using common noninvasive oxygenation devices. A simulated mannequin was designed to breathe at a minute ventilation of 20 L/min and used the following oxygen-therapy devices: nasal cannula oxygenation (NCO) at 4 and 15 L/min, nonrebreathing mask (NRM) at 15 L/min, simple mask at 6 L/min, combination of NCO at 15 L/min and NRM at 15 L/min, high-flow nasal cannula (HFNC) at 50 L/min, and flush rate NRM. Two-dimension of the dispersion distance and the aerosol concentrations were measured at head, trunk, and foot around the mannequin for over 10 min. HFNC and flush-rate NRM yielded the longest dispersion distance and highest aerosol concentrations over the three sites of the mannequin than the other oxygenation devices and should use with caution. For flow rates of < 15 L/min, oxygenation devices with mask-like effects, such as NRM or NCO with NRM, decreased aerosol dispersion more effectively than NCO alone or a simple mask. In the upright position, the foot area exhibited the highest aerosol concentration regardless of the oxygenation device than the head-trunk areas of the mannequin. Healthcare workers should be alert even at the foot side of the patient while administering oxygenation therapy.

MeSH terms

  • Administration, Inhalation
  • Aerosols
  • Humans
  • Lung
  • Oxygen Inhalation Therapy* / methods
  • Smoke*

Substances

  • Smoke
  • Aerosols