Spine Surgery with Electronic Conductivity Device: A Prospectively Multicenter Randomized Clinical Trial and Literature Review

Orthop Surg. 2023 Nov;15(11):2872-2880. doi: 10.1111/os.13880. Epub 2023 Sep 22.

Abstract

Objective: Improving accuracy and safety of pedicle screw placement is of great clinical importance. Electronic conductivity device (ECD) can be a promising technique with features of affordability, portability, and real-time detection capabilities. This study aimed to validate the safety and effectiveness of a modified ECD.

Methods: The ECD underwent a modification where six lamps of various colors, and it was utilized in a prospectively multicenter randomized controlled clinical trial involving 96 patients across three hospitals from June 2018 to December 2018. The trial incorporated a self-control randomization with an equal distribution of left or right side of vertebral pedicle among two groups: the free-hand group and the ECD group. A total of 496 pedicle screws were inserted, with 248 inserted in each group. The primary outcomes focused on the accuracy of pedicle screw placement and the frequency of intraoperative X-rays. Meanwhile, the secondary indicator measured the time required for pedicle screw placement. Results were presented as means ± SD. Paired samples t-test and χ2 -test were used for comparison. Furthermore, an updated review was conducted, which included studies published from 2006 onwards.

Results: Baseline patient characteristics were recorded. The primary accuracy outcome revealed a 96.77% accuracy rate in the ECD group, compared to a 95.16% accuracy rate in the free-hand group, with no significant differences noted. In contrast, ECD demonstrated a significant reduction in radiation exposure frequency when compared to the free-hand group (1.11 ± 0.32 vs. 1.30 ± 0.53; p < 0.001), resulting in a 14.6% reduction. Moreover, ECD displayed a decrease of 30.38% in insertion time (70.88 ± 30.51 vs. 101.82 ± 54.00 s; p < 0.001). According to the results of the 21 studies, ECD has been utilized in various areas of the spine such as the atlas, thoracic and lumbar spine, as well as sacral 2-alar-iliac. The accuracy of ECD ranged from 85% to 100%.

Conclusion: The prospectively randomized trial and the review indicate that the use of ECD presents a secure and precise approach to the placement of pedicle screws, with the added benefit of reducing both procedure time and radiation exposure.

Keywords: Accuracy; Electronic conductivity device; Navigation surgery; Pedicle screw; Radiation exposure.

Publication types

  • Review

MeSH terms

  • Electronics
  • Humans
  • Lumbar Vertebrae / surgery
  • Multicenter Studies as Topic
  • Pedicle Screws*
  • Radiography
  • Randomized Controlled Trials as Topic
  • Sacrum
  • Spinal Fusion* / methods
  • Surgery, Computer-Assisted* / methods