Novel tongue-positioning device to reduce tongue motions during radiation therapy for head and neck cancer: Geometric and dosimetric evaluation

PLoS One. 2023 Sep 21;18(9):e0291712. doi: 10.1371/journal.pone.0291712. eCollection 2023.

Abstract

This study aimed to assess the performance of a tongue-positioning device in interfractional tongue position reproducibility by cone-beam computed tomography (CBCT). Fifty-two patients treated with radiation therapy (RT) while using a tongue positioning device were included in the study. All patients were treated with 28 or 30 fractions using the volumetric modulated arc therapy technique. CBCT images were acquired at the 1st, 7th, 11th, 15th, 19th, 23th, and 27th fractions. Tongues on planning computed tomography (pCT) and CBCT images were contoured in the treatment planning system. Geometric differences in the tongue between pCT and CBCT were assessed by the Dice similarity coefficient (DSC) and averaged Hausdorff distance (AHD). Two-dimensional in vivo measurements using radiochromic films were performed in 13 patients once a week during sessions. The planned dose distributions were compared with the measured dose distributions using gamma analysis with criteria of 3%/3 mm. In all patients, the mean DSC at the 1st fraction (pCT versus 1st CBCT) was 0.80 while the mean DSC at the 27th fraction (pCT versus 27th CBCT) was 0.77 with statistical significance (p-value = 0.015). There was no statistically significant difference in DSC between the 1st fraction and any other fraction, except for the 27th fraction. There was statistically significant difference in AHD between the 1st fraction and the 19th, 23th, and 27th fractions (p-value < 0.05). In vivo measurements showed an average gamma passing rate of 90.54%. There was no significant difference between measurements at the 1st week and those at other weeks. The tongue geometry during RT was compared between pCT and CBCT. In conclusion, the novel tongue-positioning device was found to minimize interfractional variations in position and shape of the tongue.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Head and Neck Neoplasms* / diagnostic imaging
  • Head and Neck Neoplasms* / radiotherapy
  • Humans
  • Radiometry
  • Radiotherapy, Intensity-Modulated*
  • Reproducibility of Results
  • Tongue / diagnostic imaging

Grants and funding

This study was supported by 'Supporting Project to evaluation New Domestic Medical Devices in Hospitals' funded by Ministry of Health and Welfare (MOHW) and Korea Health Industry Development Institute (KHIDI). JHK and JIK received the funding. The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.