LC-MS/MS methods for determination of venetoclax in human plasma and cerebrospinal fluid

Biomed Chromatogr. 2023 Dec;37(12):e5738. doi: 10.1002/bmc.5738. Epub 2023 Sep 18.

Abstract

We developed and validated sensitive MS/MS methods for the determination of venetoclax, an oral selective B-cell lymphoma-2 inhibitor, in human plasma and cerebrospinal fluid (CSF). Acetonitrile was used as protein precipitant. The mobile phase was 10 mM ammonium formate consisting of 0.1% formic acid and acetonitrile (40:60, v/v). The analytes were separated on an ACQUITY UPLC HSS T3 column (2.1 × 50 mm, 1.8 μm) in 5 min. An API 4000 mass spectrometer was selected to quantify venetoclax and internal standard using m/z 868.3 → 636.3 and 876.3 → 644.3 under multiple response monitoring mode. In plasma, the calibration curve exhibited good linearity ranging from 20.0 to 5000 ng/mL, whereas in the CSF, the linear range was 0.500-100 ng/mL. The matrix effect of venetoclax and internal standard (venetoclax-d8) was not obvious in both plasma and CSF. The inter- and intra-run accuracy was within ±11.9%, and the inter- and intra-run precision was below 13.6%. Both methods had no carryover, and the recovery was close to 100%. The validated methods were employed to quantify the concentrations of venetoclax in the plasma and CSF of patients diagnosed with chronic lymphocytic leukemia or acute myelogenous leukemia.

Keywords: MS/MS; cerebrospinal fluid; human plasma; venetoclax.