Interlaboratory comparison of electron paramagnetic resonance tooth enamel dosimetry with investigations of the dose responses of the standard samples

Radiat Prot Dosimetry. 2023 Sep 18;199(14):1557-1564. doi: 10.1093/rpd/ncad150.

Abstract

A total of seven Japanese laboratories participated in an intercomparison study to estimate the dose given to tooth enamel samples, using the electron spin resonance method. Each of four of the participating laboratories prepared a set of tooth enamel samples, using the electron spin resonance method. Four of the participating laboratories each prepared a set of tooth enamel samples, consisting of seven standard aliquots irradiated from 100 to 2000 mGy and three samples with an 'unknown' dose between 140 and 960 mGy, were intended to eliminate bias from sample preparation. Although not all seven laboratories measured all four sets of samples, the major finding was that systematic biases in estimating doses may be caused by differences in laboratory measurements rather than by the enamel extracting procedures. When doses were averaged by measurements made by multiple laboratories, the averaged values were close to the actual values. Scattering in the intercepts in the standard dose response would be a serious problem in actual dosimetry where no background sample is available.