We compared the performance of the STANDARD F and SD BIOLINE stool antigen tests in 335 patients. The performance of STANDARD F (sensitivity: 95.6%; specificity: 94%) was highly comparable to that of SD BIOLINE (sensitivity: 92.6%; specificity: 93.5%), suggesting that STANDARD F is useful for the detection of Helicobacter pylori infection.
Keywords: Helicobacter pylori; Performance; SD BIOLINE; STANDARD F; Stool antigen test.
Copyright © 2023 Elsevier Inc. All rights reserved.