Validation of the iHealth Track upper-arm blood pressure monitor KN-550BT in general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020)

Blood Press Monit. 2024 Feb 1;29(1):41-44. doi: 10.1097/MBP.0000000000000678. Epub 2023 Sep 14.

Abstract

Objective: To validate the iHealth Track KN-550BT oscillometric upper-arm blood pressure monitor in general population according to the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020).

Methods: Participants were recruited and the same left-arm sequential method was used for blood pressure measurement according to the ISO 81060-2:2018+AMD1:2020. The validation results were assessed following the protocol and the Bland-Altman scatterplot was used to show the difference between the test device and reference results.

Results: A total of 89 qualified participants were included in the final analysis. For the validation Criterion 1, the mean ± SD of the differences between the test device and reference readings was -1.22 ± 5.76 mmHg and -0.08 ± 4.40 mmHg for systolic and diastolic blood pressure, respectively. For Criterion 2, the mean ± SD of the differences between the test device and reference readings per participant was -1.22 ± 5.06 mmHg and -0.08 ± 3.84 mmHg for systolic and diastolic blood pressure, respectively.

Conclusion: The iHealth Track KN-550BT upper-arm blood pressure monitor passed all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2:2018+AMD1:2020) and can be recommended for clinical use and self-measurement in general population.