Abstract

Ultra-high frequency (>100 MHz) acoustic waves feature biocompatibility and high sensitivity and allow biomedical imaging and acoustic tweezers. Primarily, excellent spatial resolution and broad bandwidth at ultra-high frequency is the goal for pathological research and cell selection at the cellular level. Here, we propose an efficient approach to visualize mouse brain atrophy by self-focused ultrasonic sensors at ultra-high frequency with ultra-broad bandwidth. The numerical models of geometry and theoretically predicted acoustic parameters for half-concave piezoelectric elements are calculated by the differential method, which agrees with measured results (lateral resolution: 24 μm, and bandwidth: 115% at -6 dB). Compared with the brain slices of 2-month-old mouse, the atrophy visualization of the 6-month-old mouse brain was realized by C-mode imaging with an acoustic microscopy system, which is a potential prospect for diagnosis and treatment of Alzheimer's disease (AD) combined with neuroscience. Meanwhile, the acoustic properties of the brain slices were quantitatively measured by the acoustic microscopy. These encouraging results demonstrate the promising application for high-resolution imaging in vitro biological tissue with ultra-high frequency self-focusing ultrasonic sensors.