Influence of tube and patient positioning in thoracoabdominal CT examinations on radiation exposure-towards a better patient positioning

J Radiol Prot. 2023 Sep 11;43(3). doi: 10.1088/1361-6498/acf384.

Abstract

Although iso-centric patient positioning is enormously important in computed tomography (CT), it is complicated in thoracoabdominal imaging by the varying dimensions of the body. Patient positioning can affect the appearance of the patient on the localiser. Positioned too close to the x-ray tube, a patient appears considerably more voluminous. The goal of this study is to assess the difference in radiation exposure of combined chest and abdomen CT scans between scans with prior 0°- and 180°-localisers in conjunction with patient positioning. In this IRB-approved retrospective study, patients who had two routine thoracoabdominal CT scans on the same CT scanner, one with a prior 0°- and one with a prior 180°-localiser, were included. To evaluate the radiation exposure of the thoracoabdominal CT examination regarding the tube position during the localiser, volumetric computed tomography dose index (CTDIvol), size-specific dose estimate (SSDE), patient diameter and positioning within the iso-centre for three positions (heart, abdomen, femur level) were compared with regard to the tube position during the prior localiser. CT examinations of 114 patients were included. Despite similar patient weight and diameter between the two examinations, SSDE and CTDIvolwas significantly larger (up to 73%) with 180°-localisers. Patient offset from the iso-centre ranged between -9 mm at the centre slice (abdomen level) to -43 mm at the most caudal slice at the pelvis (femur level), causing a significant magnification (p < 0.001) on 180°-localisers with a subsequent increase of the apparent attenuation. The results of this study emphasise the use of 0°-localisers in thoracoabdominal CTs, since 180°-localisers caused patient magnification with subsequent increase in radiation exposure. The advantage of 180°-localisers, namely reducing the dose in thyroid and breast, is eliminated if the dose of the CT scan increases significantly in the abdomen and pelvis.

Keywords: computed tomography; localiser; patient positioning; radiation exposure.

MeSH terms

  • Humans
  • Patient Positioning
  • Radiation Exposure*
  • Retrospective Studies
  • Tomography Scanners, X-Ray Computed
  • Tomography, X-Ray Computed*