Fully Closed-Loop Glucose Control Compared With Insulin Pump Therapy With Continuous Glucose Monitoring in Adults With Type 1 Diabetes and Suboptimal Glycemic Control: A Single-Center, Randomized, Crossover Study

Diabetes Care. 2023 Nov 1;46(11):1916-1922. doi: 10.2337/dc23-0728.

Abstract

Objective: We evaluated the safety and efficacy of fully closed-loop with ultrarapid insulin lispro in adults with type 1 diabetes and suboptimal glycemic control compared with insulin pump therapy with continuous glucose monitoring (CGM).

Research design and methods: This single-center, randomized, crossover study enrolled 26 adults with type 1 diabetes using insulin pump therapy with suboptimal glycemic control (mean ± SD, age 41 ± 12 years, HbA1c 9.2 ± 1.1% [77 ± 12 mmol/mol]). Participants underwent two 8-week periods of unrestricted living to compare fully closed-loop with ultrarapid insulin lispro (CamAPS HX system) with insulin pump therapy with CGM in random order.

Results: In an intention-to-treat analysis, the proportion of time glucose was in range (primary end point 3.9-10.0 mmol/L) was higher during closed-loop than during pump with CGM (mean ± SD 50.0 ± 9.6% vs. 36.2 ± 12.2%, mean difference 13.2 percentage points [95% CI 9.5, 16.9], P < 0.001). Time with glucose >10.0 mmol/L and mean glucose were lower during closed-loop than during pump with CGM (mean ± SD time >10.0 mmol/L: 49.0 ± 9.9 vs. 62.9 ± 12.6%, mean difference -13.3 percentage points [95% CI -17.2, -9.5], P < 0.001; mean ± SD glucose 10.7 ± 1.1 vs. 12.0 ± 1.6 mmol/L, mean difference -1.2 mmol/L [95% CI -1.8, -0.7], P < 0.001). The proportion of time with glucose <3.9 mmol/L was similar between periods (median [interquartile range (IQR)] closed-loop 0.88% [0.51-1.55] vs. pump with CGM 0.64% [0.28-1.10], P = 0.102). Total daily insulin requirements did not differ (median [IQR] closed-loop 51.9 units/day [35.7-91.2] vs. pump with CGM 50.7 units/day [34.0-70.0], P = 0.704). No severe hypoglycemia or ketoacidosis occurred.

Conclusions: Fully closed-loop insulin delivery with CamAPS HX improved glucose control compared with insulin pump therapy with CGM in adults with type 1 diabetes and suboptimal glycemic control.

Trial registration: ClinicalTrials.gov NCT04977908.

Publication types

  • Randomized Controlled Trial
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Blood Glucose
  • Blood Glucose Self-Monitoring
  • Cross-Over Studies
  • Diabetes Mellitus, Type 1* / drug therapy
  • Humans
  • Hypoglycemic Agents / therapeutic use
  • Insulin / therapeutic use
  • Insulin Infusion Systems
  • Insulin Lispro / therapeutic use
  • Insulin, Regular, Human / therapeutic use
  • Middle Aged
  • Treatment Outcome

Substances

  • Hypoglycemic Agents
  • Insulin
  • Blood Glucose
  • Insulin Lispro
  • Insulin, Regular, Human

Associated data

  • ClinicalTrials.gov/NCT04977908