Early effects of left bundle branch area pacing on ventricular activation by speckle tracking echocardiography

J Interv Card Electrophysiol. 2024 Mar;67(2):341-351. doi: 10.1007/s10840-023-01616-7. Epub 2023 Aug 21.

Abstract

Background: Left bundle branch area pacing (LBBAP) is an emerging cardiac pacing modality that preserves fast electrical activation of the ventricles and provides very good electrical measures. Little is known on mechanical ventricular activation during this pacing modality.

Methods: We prospectively enrolled patients receiving LBBAP. Electrocardiographic and electrical parameters were evaluated at implantation, < 24 h and 3 months. Transthoracic echocardiography with strain analysis was performed at baseline and after 3 months, when ventricular mechanical activation and synchrony were analyzed by time-to-peak standard deviation (TPSD) of strain curves for both ventricles. Intraventricular left ventricular (LV) dyssynchrony was investigated by LV TPSD and interventricular dyssynchrony by left ventricle-right ventricle TPSD (LV-RV TPSD).

Results: We screened 58 patients with permanent pacing indication who attempted LBBAP. Procedural success was obtained in 56 patients (97%). Strain data were available in 50 patients. QRS duration was 124.1 ± 30.7 ms at baseline, while paced QRS duration was 107.7 ± 13.6 ms (p < 0.001). At 3 months after LBBAP, left ventricular ejection fraction (LVEF) increased from 52.9 ± 10.6% at baseline to 56.9 ± 8.4% (p = 0.004) and both intraventricular LV dyssynchrony and interventricular dyssynchrony significantly improved (LV TPSD reduction from 38.2 (13.6-53.9) to 15.1 (8.3-31.5), p < 0.001; LV-RV TPSD from 27.9 (10.2-41.5) to 13.9 (4.3-28.7), p = 0.001). Ameliorations with LBBAP were consistent in all subgroups, irrespective of baseline QRS duration, types of intraventricular conduction abnormalities, and LVEF.

Conclusions: Echocardiographic strain analysis shows that LBBAP determines a fast and synchronous biventricular contraction with a stereotype mechanical activation, regardless of baseline QRS duration, pattern, and LV function.

Keywords: Conduction system pacing; Left bundle branch area pacing; Speckle tracking echocardiography; Ventricular synchrony.