A novel inductively coupled capacitor wireless sensor system for rapid antibiotic susceptibility testing

J Biol Eng. 2023 Aug 18;17(1):54. doi: 10.1186/s13036-023-00373-5.

Abstract

Background: The increasing prevalence and severity of antimicrobial resistance (AMR) present a major challenge to our healthcare system. Rapid detection of AMR is essential for lifesaving under emergent conditions such as sepsis. The current gold standard phenotypic antibiotic susceptibility testing (AST) takes more than a day to obtain results. Genotypic ASTs are faster (hours) in detecting the presence of resistance genes but require specific probes/knowledge of each AMR gene and do not provide specific information at the phenotype level. To address this unmet challenge, we developed a new rapid phenotypic AST.

Result: We designed a new electrochemical biosensor based on the concept of magnetically coupled LC sensors. The engineered LC sensors can be placed in 96-well plates and communicate the reading remotely with a receiver coil for signal analysis. The sensors were validated by monitoring the growth of Escherichia coli, Staphylococcus aureus, and Pseudomonas aeruginosa in the presence and absence of different antibiotics. Drug-resistant strains were used as controls. Bacterial growth was detected within 30 min after inoculation, allowing rapid determination of antibiotic susceptibility at the phenotype level. The sensor also functions in the presence of host proteins when tested with 2% FBS in growth media.

Conclusions: With the compatibility with 96-well plates, this label-free rapid 30-min AST has the potential for low-cost applications with simple integration into the existing workflow in clinical settings.

Keywords: Antibiotic resistance; Antimicrobial susceptibility test; Biosensor; Permittivity.